Item 1. Business

Overview

EBR is a U.S. based medical device company that developed the WiSE CRT System (“WiSE”), an implantable cardiac pacing system able to provide stimulation to endocardial heart tissue for the correction of heart rhythm conditions without requiring the use of leads. That implantable device is part of a cardiac resynchronization therapy (“CRT”), offering endocardial heart tissue stimulation without the complications associated with traditional lead-based systems. Cardiac rhythm management (“CRM”) systems use leads to conduct electricity from an implantable pulse generator (“IPG”) to electrodes that deliver therapeutic electric pulses to heart tissue. While leads are a critical part of most CRM systems, they have long been recognized as a primary shortcoming of these systems and are a leading cause of device failure.

We initially developed WiSE for use in conjunction with another implanted pacemaker to provide CRT to patients who are unable to receive CRT from a traditional lead-based system or are at high risk of complications from an upgrade procedure. WiSE CRT technology is engineered to benefit patients who have not seen success with conventional CRT or face high complication risks. By eliminating lead requirements for left ventricular pacing, WiSE CRT introduces a novel approach to cardiac pacing, with the potential to transform CRT delivery.

On April 11, 2025, we received notification that the Center for Devices and Radiological Health (“CDRH”) of the Food and Drug Administration (“FDA”) had completed its review of our premarket approval application (“PMA”) for WiSE and approved WiSE for commercial distribution in the U.S. for adult patients who are at least 22 years of age, are indicated for CRT, have an existing or are eligible for an implanted right ventricular pacing system, and are in one of the following two categories: 1) patients in whom previous coronary sinus (“CS”) lead implantation was unsuccessful, or where an implanted lead has been turned off, referred to as "previously untreatable"; or 2) patients with previously implanted pacemakers or Implantable Cardioverter-Defibrillators (“ICDs”) in whom standard CRT upgrade is not advisable due to known relative contraindications for CS lead or CRT device implantation, referred to as "high risk upgrades".

We have launched WiSE with the focus on driving adoption of WiSE at key, high-volume, hospitals or medical facilities within the U.S. to be followed by select, high-volume hospitals or medical facilities in markets outside the U.S. (“OUS”) that we would target after evaluating regulatory and reimbursement considerations. The growth of our business depends on our ability to successfully commercialize WiSE and gain wide acceptance of WiSE by continuing to make physicians and other hospital staff aware of the benefits of WiSE to generate increased demand and frequency of use and thus increase sales to our hospital customers. Our ability to grow our business will also depend on our ability to expand our customer base in existing or new target markets. The rate at which we grow our sales force and the speed at which newly hired salespeople become effective can impact our revenue growth or our costs incurred in anticipation of such growth. We intend to make significant investments in our sales and marketing organization by increasing the number of U.S. sales representatives and expanding our international marketing programs to help facilitate further adoption among existing hospital accounts as well as broaden awareness of our products to new hospital accounts.

Products

The WiSE CRT System, our only product, is an implantable cardiac pacing system capable of delivering pacing level energy to the heart without using a lead/ wire. The technology used to achieve this leadless pacing is based on converting ultrasound energy into electrical energy. The system consists of a battery connected to an ultrasound transmitter that is implanted subcutaneously and the electrode implanted in the LV endocardium. The system requires a co-implant (e.g., pacemaker, defibrillator, or CRT) capable of right ventricular (“RV”) pacing. The transmitter senses the RV pacing spike of the co-implant and within approximately five milliseconds emits an ultrasonic pulse to the electrode, which converts the ultrasound energy received to electrical energy at sufficient amplitudes to pace/ stimulate cardiac tissue. The intensity of the ultrasound energy used is very low, even in comparison to levels used for echocardiographic imaging.

The leadless WiSE Electrode essentially replaces the pacing function of a traditional LV lead. The WiSE CRT System is used in conjunction with a typical, commercially available implanted pacemaker, implantable cardioverter defibrillator (“ICD”), or CRT device with an RV pacing function. Immediately after sensing an RV pacing output from the co-implanted device, the WiSE CRT System triggers an ultrasound pulse targeted at the Electrode to pace the left ventricle. The sequence of sensing, transmitting, receiving, and stimulating the left ventricle is essentially simultaneous with the co-implanted device’s RV pacing output and thus provides biventricular (“BiV”) pacing analogous to CRT pacing devices.

Components of the WiSE CRT System include the Delivery Sheath and Electrode Catheter which delivers the Electrode, the Pulse Generator (Transmitter and Battery), and the Programmer (Figure 1.1 below).

Figure 1.1 How WiSE CRT System Provides Leadless Cardiac Pacing

SOLVE-CRT Clinical Trial

The SOLVE-CRT (Stimulation of the Left Ventricular Endocardium for Cardiac Resynchronization Therapy) investigational device exemption (“IDE”) study was a prospective, multicenter trial consisting of an initial randomized, double-blind part and a subsequent single-arm part. It was originally designed as a randomized, multinational, double-blind study to enroll 350 patients from up to 45 centers. Patients were enrolled if they received right ventricular (“RV”) pacing from a prior pacing implant and either did not have a fully functioning CRT system because of lead issues (Previously Untreatable [“PU”]), did not respond to CRT therapy (Non-Responders [“NR”]), or were considered high risk for a standard CRT upgrade (High-Risk Upgrade [“HRU”]), which included those with a cardiac pacemaker or intracardiac defibrillator (“ICD”) requiring upgrade to CRT.

The first patient was enrolled in the SOLVE-CRT study in January 2018. Enrollment was severely impacted by the COVID-19 pandemic and was paused in March 2020 after 108 patients were enrolled. At that time, the investigators worked with the FDA to revise the clinical protocol, implementing a single-arm, non-randomized part to complete the study. Central to this strategy was a differentiation of the three original patient groups and the requirements for demonstration of safety and efficacy. For the patients in the PU and HRU groups, CRT was an approved therapy, and the WiSE CRT System could be viewed as an alternate method of providing CRT when conventional CRT was not possible, so a Single-Arm study with pre-specified objective performance goals was appropriate. In contrast, the NR group had suboptimal responses to conventional CRT due to a diverse set of pathophysiologic mechanisms. Thus, there was a different threshold for scientifically acceptable evidence for safety and effectiveness, and this group was excluded from the study continuation.

The modified study was designed to enroll up to 300 patients with a pre-specified interim analysis with early stopping rules after enrolling 183 patients. All 183 patients (PU/HRU/NR) were included in the safety analysis while only the PU and HRU patients (n = 100) were included in the efficacy analysis. The primary safety endpoint was the freedom from device- or procedure-related complications (Type I). The primary efficacy endpoint was the reduction in left ventricular systolic volume (“LVESV”). LVESV is a key marker of advanced heart failure and goes up as the HF progresses.

At the interim analysis, the primary 6-month efficacy endpoint was met with a 16.4% (95% confidence interval [“CI”], 11.7% to 21.0%) reduction in mean LVESV, significantly favorable to the 9.3% performance goal (p = 0.003). The primary 6-month safety endpoint was met with an 80.9% (148/183 participants) (lower boundary of the one-sided 98.8% CI, 73.4%) rate of freedom from device- or procedure-related (Type I) complications, significantly favorable to the 70% performance goal (p < 0.001). Since both the primary efficacy and safety endpoints met the pre-specified stopping rules for interim analysis, the trial was concluded early for success.

Secondary endpoints included an acoustic pacing capture threshold (“APCT”) of < 2.9 mJ achieved in 95.2% of participants and APCT stability achieved in 81.3% of participants, indicating that the WiSE CRT System was capable of delivering CRT in the majority of patients enrolled in the study and that the energy required to deliver CRT remained stable. In addition, 93.1% of participants achieved a mean percent BiV pacing, 46.1% achieved an increase in Left ventricular ejection fraction (“LVEF”) ≥ 5%, and 65.5% achieved an increase of ≥ 5 points in Kansas City Cardiomyopathy Questionnaire.

The SOLVE-CRT study demonstrated that leadless left ventricular endocardial pacing (“LVEP”) with the novel WiSE CRT System is feasible, safe, and effective for delivering CRT in patients with advanced HF.

FDA Approval

On April 11, 2025, the Center for Devices and Radiological Health of the FDA completed its review of and approved our premarket approval application for the WiSE CRT System. The WiSE CRT System is indicated for adult patients who are at least 22 years of age, are indicated for cardiac resynchronization therapy, have an existing or are eligible for an implanted right ventricular pacing system and are in one of the following two categories: (a) patients in whom previous coronary sinus lead implantation was unsuccessful, or where an implanted lead has been turned off, referred to as “previously untreatable”; and (b) patients with previously implanted pacemakers or Implantable Cardioverter-Defibrillators in whom standard CRT upgrade is not advisable due to know relative contraindications for CS lead or CRT device implantation, referred to as “high risk upgrades”.

Post-approval Study

To help assure the continued safety and effectiveness of the WiSE CRT System, the FDA has required a post-approval study (“PAS”) as a condition of approval under 21 CFR 814.82(a)(2). As part of our PMA approval, we agreed with the FDA to conduct a PAS, with a goal to enroll approximately 320 patients such that 250 implanted patients will reach the 12-month follow-up for the primary endpoint. Participants will be enrolled, implanted, and followed at one month, six months, and annually thereafter for the duration of the study through 5 years follow-up. The primary endpoint is to evaluate the safety of the WiSE CRT System by assessing the rate of device- or procedure- related serious adverse events. The study will also measure: (i) procedural success; (ii) CRT response assessment; (iii) Heart Failure Clinical Composite Score; (iv) New York Heart Association Functional Classification; (v) QRS duration; (vi) 6-minute walk test; (vii) events assessment; (viii) mortality; (ix) Patient Global Assessment; and (x) device longevity and battery performance. We are required to submit an interim PMA PAS Report every six months until subject enrollment has been completed, and annually thereafter, from the date of the PMA approval.

In the fourth quarter of 2025, we enrolled four patients in the post-approval study, known as WiSE System Utilization and Performance Registry (“WiSE-UP’). The first three procedures were performed in November 2025 at St. Bernards heart and Vascular Center by globally respected electrophysiologist Dr. Devi Nair. In December 2025, one additional patient was successfully enrolled by Dr. Dinesh Sharma, at Naples Comprehensive Health (NCH).

Reimbursement

The process for determining whether a payor will provide coverage for a product is typically separate from the process for setting the reimbursement rate that the payor will pay for the product. A payor’s decision to provide coverage for a product does not imply that an adequate reimbursement rate will be available. Some products may not be reimbursed separately but their cost may instead be bundled as part of the payment received by the provider for the procedure only. Separate reimbursement for the product itself or the treatment or procedure in which a product is used may not be available. Commercial third-party payors often rely upon Medicare coverage policy and payment limitations in setting their own reimbursement policies but also have their own methods and approval process apart from Medicare coverage and reimbursement determinations. As such, one third-party payor’s determination to provide coverage for a product does not ensure that other payors will also provide coverage for the product.

As a result of its breakthrough device designation, our WiSE CRT technology is eligible for incremental payment coverage in the U.S. for up to three years following FDA approval. The Centers for Medicare & Medicaid Services (“CMS”) approved of the New Technology Add-On Payment (“NTAP”) for WiSE. The program commenced on October 1, 2025. The NTAP is designed to bridge the financial gap between the costs of innovative technologies and the standard Medicare severity Diagnosis Related Groups (“MS-DRG” or “DRG”) payment structure in place, while encouraging early adoption of the breakthrough medical advancements used in the inpatient setting for Medicare patients. NTAP secures the maximum reimbursement rate of up to $41,145 of the cost of WiSE. This is in addition to the DRG payments, which are intended to cover the procedure and the remaining device cost. CMS has set the level based on an average selling price of $63,300, which will remain in place for three years while claims data are collected. We intend to petition CMS during the second year to move WiSE procedures to a DRG that fully covers the device add procedure, which is the usual pathway for durable reimbursement. In June 2025, we received preliminary approval for Transitional Pass-Through (“TPT”) reimbursement scheme of WiSE. Preliminary approval for the TPT reimbursement scheme commenced October 1, 2025, and is effective for three years. TPT provides hospitals with Medicare reimbursement when treating patients in an outpatient setting. In combination, these reimbursements are intended to fully cover the cost of WiSE in a hospital or outpatient setting.

Markets and Distribution

After receiving FDA approval in April 2025, we commercially launched WiSE with the focus on driving early adoption of WiSE at key, high-volume, luminary sites within the U.S. to be followed by select, high-volume sites in markets outside the U.S. (“OUS”) that we are targeting after evaluating regulatory and reimbursement considerations.

a)U.S. Strategy

We commenced an initial soft launch of WiSE at select sites in the U.S. during the second and third quarters of 2025, aimed at collecting customer feedback to refine our longer-term strategy. On October 1, 2025, we launched our Limited Market Release (“LMR”), with the following objectives:

LMR Objective 1: Initial Target Accounts

·Launch in target accounts from the SOLVE-CRT pivotal trial.
·Leverage existing relationships with trial sites to streamline patient identification and device implantation.

LMR Objective 2: Expansion and Optimization

·Strategically expand our field team to broaden our market presence into additional high-volume sites.
·Focus on optimizing the customer training programs and enhancing EBR's business operations.

LMR Objective 3: Increase Implants Per Site

·The field team will focus on increasing the number of cases per month per site by improving hospital implant workflows and familiarity with the technology.

Full Market Release

We anticipate launching our full market release in the second half of 2026, which includes the following strategic objectives:

·Expand market presence and maximize product adoption.
·Utilize refined business operations and continue scaling the field team.
·Focus on expanding into additional sites, leveraging the experience and efficiency gained from the LMR.
b)OUS Strategy

Our OUS commercial activities will not commence until we obtain regulatory approvals and certification in select, target markets. These initial target markets include Australia, the United Kingdom, and the European Union. The timing of launch in each of these OUS markets thus depends on meeting additional regulatory requirements, as well as on securing the appropriate payment coverage for WiSE in each market.

Competitive Environment

Significant advances in pacing technology have been made in pacing technology have been made over the last 50 years including: multi-chamber pacing, improved rate of responsiveness, devise size reduction, internet-based remote monitoring, and market increases in battery longevity. However, the basic system format of using an implantable pulse generator (“IPG”) connected to one or more leads to stimulate the heart muscle tissue, has remained unchanged over this time.

The most recent advance in the evolution of pacemakers has been the advent of leadless cardiac pacing systems, as most complications associated with pacemakers have been due to leads. To overcome this, leadless pacing systems have been developed in which the IPG and stimulating electrode are combined into a single unit that can be fully implanted inside the heart chamber. Advantages of this approach over lead-based systems include greater efficiency, system simplicity, and ease of implantation. However, these systems also have certain limitations, including the need to retrieve the device in future years due to battery depletion, risk of cardiac perforation and thrombus, and infection risk. Additionally, because of their size and thrombogenicity (tendency to generate and release clots that might cause heart attack or stroke) they cannot be used within the left ventricle.

The three major CRM device companies (Medtronic plc, Boston Scientific, and Abbott) have each developed leadless cardiac pacemakers that can be implanted in the right ventricle, with Abbott having an approved device that can be used in the right atrium. Leadless devices are expected to play an increasingly important role in the future pacemaker market.

Leadless Cardiac Rhythm Management Landscape

Figure 1.2: Current Leadless Pacemakers for Cardiac Pacing

Opportunity for WiSE

While the leadless pacemakers currently on the market are for bradycardia indication, it is anticipated that the entry of Abbott’s Aveir DR dual chamber device could further increase the adoption of leadless pacemakers.

It has been reported that up to 30% of patients with pacemakers develop pacing induced heart failure within four years. Thus, many of the patients implanted with leadless pacemakers may require an upgrade to CRT at a later date. The WiSE CRT System is the only device able to upgrade these patients to CRT.

A 14-patient clinical study, presented during Asia-Pacific Heart Rhythm Society meeting in 2022, demonstrated that WiSE is able to work in conjunction with Medtronic’s Micra to provide BiV pacing and an entirely leadless option for upgrading these patients. Only WiSE can provide these patients with an entirely leadless upgrade solution.

Supply of Components

EBR’s WiSE CRT System is comprised of five key components:

·an implantable endocardial electrode (Electrode);
·a catheter delivery system (used to implant the Electrode);
·a transmitter that operates the system (Transmitter);
·a battery that powers the Transmitter; and
·a programmer (device that programs the WiSE CRT System).

We source components and sub-assemblies for components from external suppliers and contract manufacturers. We require our suppliers and contract manufacturers to be compliant with the relevant quality standards and certifications required to manufacture medical device products such as the WiSE CRT System.

We conduct our own quality assessment and performance testing of components and subassemblies that we receive from our suppliers. We have developed and maintain the software that runs the WiSE CRT System which is uploaded into the transmitter and programmer. We inspect and evaluate the entire system prior to supplying it for use in patients.

Our existing suppliers have the capacity and capability to manufacture at volumes sufficient to meet commercial demand for WiSE. We will bring certain manufacturing processes in-house over time as we seek to reduce the cost of production. Many of the components or sub-assemblies used in the WiSE CRT System are custom built but use standard raw materials that can be provided by a variety of suppliers. Certain components within our sensor/transmitter and the receiver/transducer are unique to the WiSE CRT System design and functionality and would require redesign efforts if we need to change vendors arise. For instance**,** our piezo electric crystal is a single source component purchased from CTS Advanced Materials. We do not currently have a formal master supplier agreement in place with CTS as we generally procure on a purchase order basis. We work closely with our suppliers and have plans and measures in place to help ensure continuity of supply while maintaining high quality and reliability. Generally, we have been able to obtain adequate supplies of such raw materials and components. However, due to the U.S. FDA’s manufacturing requirements and those of other regulatory authorities, we may not be able to quickly establish additional or replacement sources for certain components or materials if we experience a sudden or unexpected reduction or interruption in supply and are unable to develop alternative sources.

Intellectual Property

The generation and protection of intellectual property, including patents, trade secrets, trademarks, proprietary technology, proprietary manufacturing techniques, and know-how, is of critical importance in our field and in biotechnology generally. We rely on a combination of trade secrets, patent filings and other intellectual property protections in an effort to protect our WiSE CRT System as well as related methods of use. We will be able to protect our WiSE CRT System and methods of use from unauthorized use by third parties only to the extent that our technology is effectively and diligently maintained as trade secrets or where applicable, covered by valid and enforceable patents. Our commercial success may also depend on whether we can defend our patents against third-party challenges and on operating without infringing on the intellectual property rights of others.

As of December 31, 2025, our WiSE CRT System and related technologies are covered by an extensive portfolio of patents which includes 59 granted U.S. patents, 38 granted non-U.S. patents, and 19 pending patent applications that cover different aspects of the WiSE CRT System including the technological inventions incorporated in:

·leadless cardiac pacing using ultrasound transduction;
·the sensor/transmitter;
·the receiver/stimulator electrode;
·the programmer;
·the delivery system; and
·mechanisms for detecting the location of the receiver/transducer electrode.

Our patents provide protection of the technology incorporated in the WiSE CRT System and a subset of these patents have expiry dates between 2025 and 2029. One pertinent patent related to the method of using an isotropic receiver-stimulator expired in December 2025. However, other active patents (including patents in the same family) with device claims that provide similar coverage of the isotropic receiver-stimulator concepts and those directed to the overarching system and method are not expiring until 2032. Other patents directed to the technology incorporated in the WiSE CRT System expire 2030 and later.

We continue to conduct research and development activities aimed at improving the use and performance of the WiSE CRT System and related technologies. This has resulted in the recent filing of new patent applications and may result in new inventions that potentially could provide opportunities to file additional patent applications. In addition, we have experience over many years with the WiSE CRT System and the underlying technology for providing leadless cardiac stimulation using ultrasound transduction. This has resulted in extensive know-how and trade secrets that are likely to represent significant barriers to any emerging competitors considering the development of a similar product.

Trade Secrets

We also rely on trade secrets, technical know-how and continuing innovation to develop and maintain our competitive position. We seek to protect our proprietary information with respect to our employees and collaborators by obtaining executed agreements requiring protection of our trade secrets and assignment of patents to us. Internal processes around the production of such things as the receiver electrode are complex and require extensive training and fixturing, lending credence to our technical know-how reliance.

Trade secrets are only beneficial if the trade secret can be protected, which in turn requires certain internal record keeping and security measures. Further, third parties are not precluded from practicing such trade secret methods developed on their own because there is no right to prevent others from this innovation. Trade secrets are difficult to protect and enforce and therefore provide us with only limited protection. Trade secrets must be protected within the company. Those employees and former employees with knowledge of our trade secrets must not share them with a third party. It is difficult to ensure that our trade secrets will be kept secret and not shared with a third party, such as a third-party competitor. For this and more comprehensive risks related to our intellectual property, please see “Item 1A. Risk Factors—Risks Related to Intellectual Property.”

Trademarks

We also have applied for and been awarded certain trademarks as shown in the table below. We intend to maintain and protect our trademarks from unauthorized use.

CountryTrademarkStatusReg. DateReg. No
AustraliaEBR SYSTEMSRegisteredDec 20 20212212887
AustraliaWISERegisteredNov 20 20191951115
ChinaEBR SYSTEMSRegisteredMar 28 202259571858
European UnionEBR SYSTEMSRegisteredFeb 22 202218564067
European UnionWISERegisteredFeb 5 202118169605
JapanEBR SYSTEMSRegisteredJan 24 20226503797
JapanWISERegisteredAug 11 20216427134
United KingdomEBR SYSTEMSRegisteredDec 24 2021UK00003698958
United KingdomWISERegisteredFeb 8 2021UK00003592127
United States of AmericaEBR SYSTEMSRegisteredNov 22 20226908275
United States of AmericaWICSRegisteredOct 6 20093692984
United States of AmericaWICS WIRELESS CARDIAC STIMULATIONRegisteredJan 10 20175118101
United States of AmericaWISERegisteredDec 13 20226925121
United States of AmericaWISERegisteredSep 19 20237169926

Government Regulation of Medical Devices

We intend to seek regulatory approval and certification for the WiSE CRT System in Australia, UK, and EU after initial commercialization in the U.S. We will evaluate the regulatory cost associated with securing CE Certificates of Conformity and affixing the CE Mark, which would allow it to be sold and used in countries within the E.U. and, potentially although not certainly, the U.K.

United States

Regulation of Medical Devices in the United States

The U.S. Food, Drug and Cosmetic Act (“FDCA”) classifies medical devices into one of three categories based on the risks associated with the device and the level of control necessary to provide reasonable assurance of safety and effectiveness. Class I devices are deemed to be low risk and are subject to the fewest regulatory controls. Class III devices are generally the highest risk devices and are subject to the highest level of regulatory control and premarket approval to provide reasonable assurance of the device’s safety and effectiveness.

Establishments that manufacture and/or distribute devices, including manufacturers, contract manufacturers, sterilizers, repackagers and relabelers, specification developers, initial importers, manufacturers of accessories and components sold directly to the end user, and U.S. manufacturers of export-only devices, are required to register their establishments with the FDA and provide the FDA a list of the devices that they handle at their facilities.

While most Class I and some Class II devices can be marketed without prior FDA authorization, most medical devices can be legally sold within the U.S. only if the FDA has: (i) approved a premarket approval application, or PMA, prior to marketing, generally applicable to Class III devices; or (ii) cleared the device in response to a premarket notification, or 510(k) submission, generally applicable to Class I and II devices. Some devices that have been classified as Class III are regulated pursuant to the 510(k) requirements because the FDA has not yet called for PMAs for these devices.

New PMA applications or PMA supplements are required for modification to the manufacturing process, equipment or facility, quality control procedures, sterilization, packaging, expiration date, labeling, device specifications, ingredients, materials, or design of a device that has been approved through the PMA process. PMA supplements often require submission of the same type of information as an initial PMA application, except that the supplement is limited to information needed to support any changes from the device covered by the approved PMA application and may or may not require as extensive technical or clinical data or the convening of an advisory panel, depending on the nature of the proposed change.

Post-market Requirements

After a device is placed on the market, numerous regulatory requirements apply. These include: Quality Management System Regulation (“QMSR”), labeling regulations, the FDA’s general prohibition against promoting products for unapproved or off-label uses, the Medical Device Reporting (“MDR”) regulation (which requires that manufacturers report to the FDA if their device may have caused or contributed to a death or serious injury or malfunctioned in a way that would likely cause or contribute to a death or serious injury if it were to recur), and the Reports of Corrections and Removals regulation (which requires manufacturers to report recalls and field actions to the FDA if initiated to reduce a risk to health posed by the device or to remedy a violation of the FDCA.

The FDA enforces these requirements by inspection and market surveillance. If the FDA finds a violation, it can institute a wide variety of enforcement actions, ranging from a public warning letter to more severe sanctions such as: fines, injunctions, and civil penalties; recall or seizure of products; operating restrictions, partial suspension or total shutdown of production; refusing requests for 510(k) clearance or PMA approval of new products; withdrawing 510(k) clearance or PMA approvals already granted; and criminal prosecution.

OUS

On May 26, 2021, Regulation (EU) 2017/745 on Medical Devices, entered into application, repealing and replacing both Directive 93/42/EEC concerning medical devices, and Directive 90/385/EEC concerning active implantable medical devices. The Regulation and its associated guidance documents and harmonized standards govern, among other things, device design and development, preclinical and clinical or performance testing, premarket conformity assessment, registration and listing, manufacturing, labeling, storage, claims, sales and distribution, export and import and post-market surveillance, vigilance, and market surveillance. Medical devices must comply with the General Safety and Performance Requirements, or GSPRs, set out in Annex I of the Medical Devices Regulation. Compliance with these requirements is a prerequisite to be able to affix the CE mark to devices, without which they cannot be marketed or sold in the European Economic Area (“EEA”). To demonstrate compliance with the GSPRs provided in the MDR and obtain the right to affix the CE mark, medical devices manufacturers must undergo a conformity assessment procedure, which varies according to the type of medical device and its classification. Apart from low risk medical devices (Class I with no measuring function and which are not sterile), in relation to which the manufacturer may issue an EU Declaration of Conformity based on a self-assessment of the conformity of its products with the GSPRs, a conformity assessment procedure requires the intervention of a Notified Body, which is an organization designated by a Competent Authority of an EEA country to conduct conformity assessments. Depending on the relevant conformity assessment procedure, the Notified Body will audit and examine the technical documentation and the quality system for the manufacture, design, and final inspection of the medical devices. The Notified Body issues a CE Certificate of Conformity following successful completion of a conformity assessment procedure conducted in relation to the medical device and its manufacturer and their conformity with the GSPRs. This Certificate and the related conformity assessment process entitle the manufacturer to affix the CE mark to its medical devices after having prepared and signed a related EU Declaration of Conformity.

As a general rule, demonstration of conformity of medical devices and their manufacturers with the GSPRs must be based, among other things, on the evaluation of clinical data supporting the safety and performance of the products during normal conditions of use. Specifically, a manufacturer must demonstrate that the device achieves its intended performance during normal conditions of use and that the known and foreseeable risks, and any adverse events, are minimized and acceptable when weighed against the benefits of its intended performance, and that any claims made about the performance and safety of the device (e.g., product labeling and instructions for use) are supported by suitable evidence. This assessment must be based on clinical data, which can be obtained from (1) clinical studies conducted on the devices being assessed, (2) scientific literature from similar devices whose equivalence with the assessed device can be demonstrated or (3) both clinical studies and scientific literature. The conduct of clinical studies in the EEA is governed by detailed regulatory obligations. These may include the requirement of prior authorization by the Competent Authorities of the country in which the study takes place and the requirement to obtain a positive opinion from a competent Ethics Committee. This process can be expensive and time-consuming. After a device is placed on the market in the EEA, it remains subject to significant regulatory requirements.

The changes to the regulatory system implemented in the EU by the Medical Devices Regulation include stricter requirements for clinical evidence and pre-market assessment of safety and performance, new classifications to indicate risk levels, requirements for third party testing by Notified Bodies, tightened and streamlined quality management system assessment procedures and additional requirements for the quality management system, additional requirements for traceability of products and transparency as well as a refined responsibility of economic operators, and the requirement to provide clinical data in the form of a clinical evaluation report.

The UK has devised a new route to market culminating in a United Kingdom Conformity Assessment (“UKCA”) Mark to replace the CE Mark. Northern Ireland will, however, continue to be covered by the EU regulations governing CE Marks. Medical devices are regulated in the United Kingdom under the Medical Devices Regulations 2002, as amended, which implement requirements derived from the EU Medical Devices Directive. The MHRA is responsible for the regulation of medical devices in the United Kingdom. Manufacturers must demonstrate conformity with applicable essential requirements, including analytical and clinical performance, before placing medical devices on the UK market. The UK has devised a new route to market culminating in a United Kingdom Conformity Assessment (“UKCA”) Mark to replace the CE Mark for medical devices placed on the Great Britain market. However, CE Marking continues to be recognized until June 30, 2028 or 2030, depending on the device type and the applicable EU regulatory regime. Northern Ireland remains subject to the EU Medical Devices Regulations under the terms of the Windsor Framework. Manufacturers must appoint a UK Responsible Person if they are not established in the United Kingdom. The United Kingdom is developing a new regulatory framework for medical devices based on a risk-based approach similar to the EU's Medical Devices Regulation with the stated aim of reducing regulatory burden, with UK-specific modifications. The new regime will include enhanced clinical evidence requirements, increased scrutiny by UK-approved bodies for higher-risk devices and strengthened post-market surveillance. The MHRA will provide transitional arrangements for compliance, though specific timelines remain subject to regulatory development and parliamentary approval.

Other Healthcare Laws

In the United States, we are subject to a number of federal and state healthcare regulatory laws that restrict business practices in the healthcare industry. These laws include, but are not limited to, federal and state anti-kickback laws, false claims laws, data privacy and security laws, and other healthcare fraud and abuse laws, such as transparency laws regarding payments or other items of value provided to healthcare providers.

•The Anti-Kickback Statute, which prohibits, among other things, knowingly and willingly soliciting, offering, receiving or paying remuneration, directly or indirectly, overtly or covertly, in cash or in kind, to induce or reward either the referral of an individual, or the purchase, order or recommendation of, items or services for which payment may be made, in whole or in part, under a federal healthcare program, such as the Medicare and Medicaid programs. The term “remuneration” has been broadly interpreted to include anything of value, and the government can establish a violation of the Anti- Kickback Statute without proving that a person or entity had actual knowledge of the law or a specific intent to violate. In addition, the government may assert that a claim, including items or services resulting from a violation of the Anti-Kickback Statute, constitutes a false or fraudulent claim for purposes of the federal civil False Claims Act. The Anti-Kickback Statute is subject to evolving interpretations and has been applied by government enforcement officials to a number of common business arrangements in the medical device industry. There are a number of statutory exceptions and regulatory safe harbors protecting certain business arrangements from prosecution under the Anti- Kickback Statute; however, those exceptions and safe harbors are drawn narrowly, and there may be limited or no exception or safe harbor for many common business activities, such as reimbursement support programs, educational and research grants, or charitable donations. Practices that involve remuneration to those who prescribe, purchase, or recommend medical devices, including discounts, providing items or services for free or engaging such individuals as consultants, advisors, or speakers, may be subject to scrutiny if they do not fit squarely within an exception or safe harbor and would be subject to a facts and circumstances analysis to determine compliance with the Anti-Kickback Statute.
•Federal civil and criminal false claims laws, including the federal civil False Claims Act, and civil monetary penalties laws, which prohibit, among other things, persons or entities from knowingly presenting, or causing to be presented, a false or fraudulent claim for payment of government funds and knowingly making, using or causing to be made or used, a false record or statement to get a false claim paid or to avoid, decrease or conceal an obligation to pay money to the federal government. A claim including items or services resulting from a violation of the Anti-Kickback Statute constitutes a false or fraudulent claim for purposes of the federal civil False Claims Act. Actions under the federal civil False Claims Act may be brought by the government or as a qui tam action by a private individual in the name of the government. These individuals, sometimes known as “relators” or, more commonly, as “whistleblowers,” may share in any amounts paid by the entity to the government in fines or settlement. Many pharmaceutical and medical device manufacturers have been investigated and have reached substantial financial settlements with the federal government under the federal civil False Claims Act for a variety of alleged improper activities, including causing false claims to be submitted as a result of the marketing of their products for unapproved and thus non-reimbursable uses and interactions with prescribers and other customers, including those that may have affected their billing or coding practices and submission of claims to the federal government. Federal civil False Claims Act liability is potentially significant in the healthcare industry because the statute provides for treble damages and mandatory monetary penalties for each false or fraudulent claim or statement. Because of the potential for large monetary exposure, healthcare and medical device companies often resolve allegations without admission of liability for significant and material amounts to avoid the uncertainty of treble damages and per claim penalties that may be awarded in litigation proceedings.
•Health Insurance Portability and Accountability Act (“HIPAA”), which imposes criminal and civil liability for, among other actions, knowingly and willfully executing, or attempting to execute, a scheme to defraud any healthcare benefit program, including private third-party payors, or knowingly and willfully falsifying, concealing or covering up a material fact or making a materially false, fictitious or fraudulent statement or representation, or making or using any false writing or document knowing the same to contain any materially false, fictitious or fraudulent statement or entry in connection with the delivery of or payment for healthcare benefits, items or services;
•HIPAA, as amended by the Health Information Technology for Economic and Clinical Health Act, or HITECH Act, and their implementing regulations, also impose obligations, including mandatory contractual terms, on covered entities subject to the rule, such as health plans, healthcare clearinghouses and certain healthcare providers, as well as their business associates and their subcontractors that perform certain services for them or on their behalf involving the use or disclosure of individually identifiable health information with respect to safeguarding the privacy, security and transmission of individually identifiable health information;
•The federal Physician Payments Sunshine Act, also known as Open Payments, which requires manufacturers of drugs, devices, biologics and medical supplies for which payment is available under Medicare, Medicaid or the Children’s Health Insurance Program to report annually, with certain exceptions, to the CMS information related to payments or other “transfers of value” made to physicians, (defined to include doctors, dentists, optometrists, podiatrists and chiropractors), certain other healthcare professionals (such as physician assistants and nurse practitioners), and teaching hospitals, and requires applicable manufacturers and group purchasing organizations to report annually to CMS ownership and investment interests held by physicians and their immediate family members; and
•Analogous state and foreign law equivalents of each of the above federal laws, such as anti-kickback and false claims laws which may apply to items or services reimbursed by any third-party payor, including commercial insurers; state and local laws that require certain regulatory licenses to manufacture or distribute medical devices commercially and/or the registration of sales representatives in the jurisdiction; state and foreign laws that require medical device companies to comply with the industry’s voluntary compliance guidelines and the applicable compliance guidance promulgated by the federal government or otherwise restrict payments that may be made to healthcare providers and other potential referral sources; state and foreign beneficiary inducement laws, which are laws that require medical device manufacturers to report information related to payments and other transfers of value to physicians and other healthcare providers or marketing expenditures; and state and foreign laws governing the privacy and security of health information in certain circumstances, many of which differ from each other in significant ways and may not have the same effect, thus complicating compliance efforts.

Ensuring that our internal operations and business arrangements with third parties comply with applicable healthcare laws and regulations will likely be costly. It is possible that governmental authorities will conclude that our business practices do not comply with current or future statutes, regulations or case law involving applicable fraud and abuse or other healthcare laws and regulations. If our operations were found to be in violation of any of these laws or any other governmental regulations that may apply to us, we may be subject to significant civil, criminal and administrative penalties, damages, fines, disgorgement, imprisonment, possible exclusion from government funded healthcare programs, contractual damages, reputational harm, diminished profits and future earnings, additional reporting obligations and oversight if we become subject to a corporate integrity agreement or other agreement to resolve allegations of non-compliance with these laws, and curtailment of our operations, any of which could substantially disrupt our operations. If the physicians or other providers or entities with whom we expect to do business are found not to be in compliance with applicable laws, they may be subject to significant criminal, civil or administrative sanctions, including exclusions from government funded healthcare programs.

Healthcare Reform

In the United States and certain foreign jurisdictions, there have been, and we expect there will continue to be, a number of legislative and regulatory changes to the healthcare system. For example, implementation of the Patient Protection and Affordable Care Act, as amended by the Health Care and Education Affordability Reconciliation Act (collectively, the ACA) substantially changed the way healthcare is financed by both governmental and private insurers in the United States and significantly affected the pharmaceutical industry.

Since its enactment, there have been judicial, administrative, executive, and Congressional legislative challenges and amendments to certain aspects of the ACA. For example, on July 4, 2025, the One Big Beautiful Bill Act (OBBBA) was signed into law, which narrowed access to ACA marketplace exchange enrollment and declined to extend the ACA enhanced advanced premium tax credits that expired at the end of 2025, which, among other provisions in the law, are anticipated to reduce the number of Americans with health insurance. The OBBBA also is expected to reduce Medicaid spending and enrollment by implementing work requirements for some beneficiaries, capping state-directed payments, reducing federal funding, and limiting provider taxes used to fund the program. Congress is considering proposed legislation intended to further reduce healthcare costs with alternatives to replace the expired ACA subsidies. It is unclear how such challenges and the healthcare reform measures of the Trump administration will impact the ACA.

Other legislative changes have been proposed and adopted since the ACA was enacted, including aggregate reductions of Medicare payments to providers of 2% per fiscal year pursuant to the Budget Control Act of 2011 which went into effect on April 1, 2013, and due to subsequent legislative amendments, will remain in effect until 2032, unless additional Congressional action is taken.

For example, on July 4, 2025, the One Big Beautiful Bill Act (OBBBA) was signed into law, which narrowed access to ACA marketplace exchange enrollment and declined to extend the ACA enhanced advanced premium tax credits that expired at the end of 2025, which, among other provisions in the law, are anticipated to reduce the number of Americans with health insurance. The OBBBA also is expected to reduce Medicaid spending and enrollment by implementing work requirements for some beneficiaries, capping state-directed payments, reducing federal funding, and limiting provider taxes used to fund the program. Congress is considering proposed legislation intended to further reduce healthcare costs with alternatives to replace the expired ACA subsidies.

Individual states in the United States have also become increasingly active in implementing regulations designed to control pharmaceutical and medical device product pricing, including price or patient reimbursement constraints, discounts, restrictions on certain product access and marketing cost disclosure and transparency measures and, in some cases, mechanisms to encourage importation from other countries and bulk purchasing. In addition, regional healthcare authorities and individual hospitals are increasingly using bidding procedures to determine which products and supplies will be included in their healthcare programs. Furthermore, there has been increased interest by third-party payors and governmental authorities in reference pricing systems and publication of discounts and list prices.

In the EU, some EU Member States may, after a medical device is CE marked, require the completion of additional studies that compare the cost-effectiveness of a particular medical device candidate to currently available therapies. This Health Technology Assessment (HTA), process is the procedure according to which the assessment of the public health impact, therapeutic impact and the economic and societal impact of use of a given medical device in the national healthcare systems of the individual country is conducted. The outcome of HTA regarding specific medical devices will often influence the pricing and reimbursement status granted to these products by the competent authorities of individual EU Member States. At the EU level, on January 12, 2025, Regulation No 2021/2282 on Health Technology Assessment (HTA Regulation), entered into application through a phased implementation. Select high-risk medical devices came into scope in 2026. The HTA Regulation is intended to boost cooperation among Member States in assessing health technologies, including newmedical devices. The Regulation establishes a framework for EU-level joint clinical assessments and increased cooperation among Member States on clinical aspects of health technology evaluation. Individual EEA countries will continue to be responsible for assessing non-clinical (e.g., economic, social, ethical) aspects of health technologies, and making decisions on pricing and reimbursement.

Jurisdiction of Incorporation

We are incorporated in the State of Delaware, United States of America, and are a registered foreign company in Australia. As a foreign company registered in Australia, we are subject to different reporting and regulatory regimes than Australian companies. As a foreign company registered in Australia, we are not subject to Chapters 6, 6A, 6B and 6C of the Corporations Act 2001 (Cth) of Australia (“Corporations Act”) dealing with the acquisition of shares (including substantial shareholdings and takeovers).

Employees

As of December 31, 2025, we had 135 full-time employees including 90 in research and development and 45 in selling, general and administrative. As of December 31, 2025, we had 126 employees located in the U.S., six in Europe, and three in Australia. None of our employees are represented by a labor union or covered by a collective bargaining agreement. We consider our relations with our employees to be good.

Available Information

Copies of our Annual Reports on Form 10-K, Quarterly Reports on Form 10-Q, Current Reports on Form 8-K, proxy statements as well as any amendments to those reports, are filed or furnished with the Securities and Exchange Commission (the “SEC”). Such reports and other information filed by the Company with the SEC are available on our website at https://www.ebrsystemsinc.com/ as soon as reasonably practicable after it is electronically filed with the SEC. References to our website address do not constitute incorporation by reference of the information contained on the website, and the information contained on the website is not part of this document. The SEC maintains an Internet site that contains reports, proxy and information statements and other information regarding issuers that file electronically with the SEC at www.sec.gov.